Equipment type:Fully automatic discrete, STAT priority
Analysis rate:Constant speed 200T/H
Test principle:Colorimetry, turbidimetry
Analysis method:End-point, 2 kinetics, fixed-time, etc. support single/double wavelength, linear and non-linear calibration
Sample System
Sample Tray:60 sample positions, random access of all samples
Sample Volume:2-50ul, 0.1ul increment
Sample probe:Liquid level detection, the remaining detection, collision protection etc.
Sample Containers:Sample cups, Micro tube, Original blood-collection tube, etc.
Reagent System
Reagent Tray:45 reagent positions
Reagent Volume:R1:25-300ul, R2:10-150ul, 0.1ul increment
Reagent Probe:Liquid level detection, the remaining detection, collision protection etc.
Reagent Refrigeration:4-15ºC, independent power supply, 24-hour non-stop refrigeration
Reaction System
Reaction Tray:Rotating tray, 120 cuvettes
Cuvette:Reusable high quality cuvette, Optical diameter 6mm
Reaction Volume:180-500ul
Incubation system:Bi-directional liquid-air device
Temperature:37ºC±0.1ºC with real-time display
Optical System
Light source:12V35W halogen lamps, over 2000 hours
Photoelectron road:after spectrophotometry
Wavelength:340,405,450,510,546,578,630,700nm
Absorbance range:0-3.500Abs
Resolution 0.0001Abs
Weight
|
80KG
|
Dimension
|
Bench top: 850mm*650mm*690mm
|
Water consumption
|
6L/Hour
|
Humidity
|
35-80%
|
Ambient Temperature
|
15-30 ºC
|
Power supply
|
AC 220V±22V, 50Hz±1Hz, power 1.5KVA
|
System connection
|
RS-232
|
PC operating system
|
Windows XP or above
|
Analysis control software
|
English version graphical operating software
|
Resolution
|