Customization: | Available |
---|---|
Classification: | Biological Diagnostics |
Type: | Biochemistry Analyzer |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Parameter:
Type: Fully automated, desktop, random access, clinical chemistry analyzer
Throughput: 130 tests / hour, up to 200 tests / hour with ISE (optional)
Test Mode: End point, Fixed time (two point), Kinetics (velocity), Immuno assays; Monochromatic, Bi-chromatic & Multi-chromatic; Linearity & Nonlinearity,Single / Dual Reagents, Monochromatic / Bi-chromatic testing.
Test Range: Clinical Biochemistry, Specific protein, ISE, Therapeutic medicine, Drug abuse, Urine/ Cerebrospinal fluid, etc.
Sample tray: 28 sample positions, any position can be aligned as STAT which can be inserted at any time during the routine examination.
Sample tubes: Sample cup, Vacuum blood collection tube, plastic tubes are all available
Sample volume: 2μl~35μl, programmed in 0.1μl stepReagent tray: 56 reagent positions for R1 and R2
Refrigeration: 24 hours non-stop cooling with Peltier element maintaining temperature 4~15?C; controlled by a separately
switch.
Reagent bottle: 15ml, 35ml and 40ml; slope bottom designReagent volume: 150μl ~300μl, programmed in 1μl step, reagent blank can be deducted automatically when testing.
Sample & Reagent Probe: Integrated multi-function reagent needle with liquid surface detection, collision protection, reagent pre-heating function, reagent residual volume monitoring and alarming function .
Probe Cleaning: inner and outer wall auto washing with warm water, carry over < 0.1%
Syringe: American imported , long-life span, high-accuracy, durable and maintenance free
Optional item: Bar code readerReaction tray: 48 reaction cuvettes, 6mm optical path, plastic cuvette, semi-permanent
Reaction volume: 150μl ~300μl
Reaction time: about 10 minutes
Temperature: 37°C, fluctuation + 0.1°C
Light source: Halogen lamp, 6V/10W, life span ≥ 3000 hours
Wavelength: 340nm, 405nm, 450nm, 492nm, 505nm, 546nm, 578nm, 620nm, 670nm,700nm
Absorbance range: 0.000 Abs ~3.000 Abs
Resolution: 0.001Abs
Calibration & Quality Control
Calibration mode: Linear (one-point, two-point and multi-point),Logit-Log4P, Logit-Log5P, Spline curve, Exponential
function, Polynomial, Parabola
QC rules: Westgard Multi-rule, Cumulative Sum Check, Twin Plot, 3 concentration levels QC
QC mode: Real time, Intra day & Inter day QC