Customization: | Available |
---|---|
Classification: | Biological Diagnostics |
Type: | Biochemistry Analyzer |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Test Items
PT, APTT, TT, Fib, RepT, VT
AT-III, Protein-S, Protein-C, LA, Hep
Standard Configuration including:
Main Unit
Test method: Nephelometry
Test channel: 1 testing channel
Heating Position: 5 positions for sample and 2 for reagent
One sample can be tested at a time
Auto pipette the measurement through configured pipettor
Single point, multi-point calibration, auto store and match with calibration curve
"CalcuLative item: S, %, PTR, INR, E.
FIB can be calculated through PT"
Net weight: 3KG
Test Method: Dispersion light nephelometric analysis
Test Principle: The highest point on first order differentiation, Percentage analysis
Zero point correction: auto track the zero point, eliminate the interference of sample and reagent
Wavelength: 470nm
Light Source: Cold Lamp with high brightness and long lifetime
Detector: unique Dispersion light detector with high sensitivity and repeatability, immunized the effect
on result of jaundice,hemolysis,chyle and turbidity on plasma
Test Channel: each channel is independent, one or more,same or different items can be tested
Simultaneously
Incubator: Provide reagent and sample with stable incubator
Incubator Temperature: 37+/-0.2°C
Measure trigger: Auto trigger the measurement through configured pipette
Item modify: easy to add or edit items, easy to set the items parameters
Calibration: single-point calibration; multipoint calibration; auto store and match with calibration curve
Quality control: auto draw the quality figure
Reagent: Open reagent system
Reagent volume: ≤40μL
Sample volume: ≤40μL
Memory: Save more than 10000 samples results, and auto recovering the results when power is off during working .
Repeatability: CV≤5%
Calculative item: S,%,PTR,INR,Fib can be get through PT, This leads to lower the cost of reagent
Printer: Inner thermal printer,can be connected to external printer
Information edit: easy to input and edit the detail information of patients
Interface: RS-232 serial port,can be connected to external printer and computer
Working Environment: Temperature:10°C~30°C;Humidity:≤70%RH
Voltage: 85-264V,Auto adapt according to the input voltage
Power : ≤80W
Quality System: Recognized by CE, TUV and CMD.
Standard Configuration including:
Main Unit
Test method: Dispersion light nephelometric analysis
Test channel: 2 testing channel
Heating well: 12 for sample, 5 for reagent
Each channel is independent ,one or two, same or different can be tested simutaneously
Auto pipette the measurement through configured pipette
Single point ,multipoint calibration, auto store and match with calibration curve
"Calculative item: S , % , PTR , INR ,E
FIB can get result through PT"
Net weight: 5KG.
Test Items: PT, APTT, TT, Fib, RepT, VT, AT-III, Protein-S, Protein-C, LA, Hep
Test Method: Dispersion light nephelometric analysis
Test Principle: The highest point on first order differentiation, Percentage analysis
Zero point correction: auto track the zero point, eliminate the interference of sample and reagent
Wavelength: 470nm
Light Source: Cold Lamp with high brightness and long lifetime
Detector: unique Dispersion light detector with high sensitivity and repeatability, immunized the effect on result of jaundice,hemolysis,chyle and turbidity on plasma
Test Channel: each channel is independent, one or more,same or different items can be tested.
Simultaneously
Incubator: Provide reagent and sample with stable incubator
Incubator Temperature: 37+/-0.2°C
Measure trigger: Auto trigger the measurement through configured pipette
Item modify: easy to add or edit items, easy to set the items parameters
Calibration: single-point calibration; multipoint calibration; auto store and match with calibration curve
Quality control: auto draw the quality figure
Reagent: Open reagent system
Reagent volume: ≤40μL
Sample volume: ≤40μL
Memory: Save more than 10000 samples results, and auto recovering the results when power is off during working .
Repeatability: CV≤5%
Calculative item: S,%,PTR,INR,Fib can be get through PT, This leads to lower the cost of reagent
Printer: Inner thermal printer,can be connected to external printer
Information edit: easy to input and edit the detail information of patients
Interface: RS-232 serial port,can be connected to external printer and computer
Working Environment: Temperature:10°C~30°C;Humidity:≤70%RH
Voltage: 85-264V,Auto adapt according to the input voltage
Power : ≤80W
Quality System: Recognized by CE, TUV and CMD.
Standard Configuration including:
Main Unit
Test method: Dispersion light nephelometric analysis
Test channel: 4 testing channels
Heating well: 24 for sample, 6 for reagent
Each channel is independent ,one or four, same or different can be tested simutaneously
Auto pipette the measurement through configured pipette
Single point ,multipoint calibration, auto store and match with calibration curve
Calculative item: S , % , PTR , INR , FIB can be get result through PT
Weight: 8KG.
Specifications:
Test Principle: The highest point on first order differentiation, Percentage analysis
Zero point correction: Auto track the zero point, eliminate the effect of sample and reagent
Wavelength: 470nm
Light Source: Cold Lamp with high brightness and long lifetime
Detector: Unique dispersion light detector with high sensitivity and repeatability, immunized the effect
on result of jaundice,hemolysis,chyle and turbidity on plasma
Test Channel: 4 channels,each channel is independent, one or more same or different items can be test
Simultaneously
Incubator: Provide 28 incubators for reagent and sample
IncubatorTemperature:37±0.2ºC
Measure trigger: Auto trigger the measurement through configured pipette
Item modify: easy to add or edit items, easy to set the items parameters
Calibration: single-point calibration; multipoint calibration; auto store and match with calibration curve
Quality control: Auto draw the QC figure
Reagent: Open reagent system
Reagent volume: ≤40μL
Sample volume: ≤40μL
Memory: Save more than 10000 samples results, and auto recover the results when power is off during
Working condition
Repeatability: CV≤5%
Calculative item: S,%,PTR,INR,Fib can be get through PT, This leads to lower the cost of reagent
Printer: Inner thermal printer,can be connected to external printer
Information edit: easy to input and edit the detail information of patients
Interface: RS-232 serial port,can be connected to external printer and computer
Working Environment: Temperature:10ºC~30ºC;Humidity:20%RH-80%RH
Voltage: AC 85V~264V,Auto adapt according to the input voltage
Power : ≤80W
Quality System: Recognized by CE TUV and CMD