Customization: | Available |
---|---|
Certification: | CE |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
1. Three-step defibrillation process
2. Two-button operation
3. Extensive voice and visual prompts for the operator
4. Biphasic energy output
5. Lock-out protection to prevent inadvertent defibrillation
6. Continuous event recording for reporting each use to a printer or computer
7. Weekly self-test to ensure readiness
Defibrillator |
|
Output |
Biphasic Truncated Exponential |
Energy Sequence |
150, 150, 200J |
Charge Time |
8 sec. to 150J; 10 sec. to 200J (New Battery) |
Analysis Time |
9 sec. |
The maximum time from the initiation of rhythm analysis to readiness for discharge with a new battery |
Less than 30 seconds |
The maximum time from the initiation of rhythm analysis to readiness for discharge after 6 shocks |
Less than 35 seconds |
The maximum time from initially switching power on to readiness for discharge |
Less than 25 seconds |
Audible Prompts |
22 audible prompts |
Visual Prompts |
LED prompts |
Controls |
Two buttons-ON/Off, Shock |
Output Energy Accuracy |
±15% into any impedance from 25 to 175Ω |
The Maximum Voltage |
1200±50V |
Output disabled when the patient impedance is outside limits |
20Ωto 200Ω |
Battery |
|
Part No. |
CR123A-4*2 |
Non-rechargeable |
12V DC, 2.8Ah |
Capacity |
At 150 Joules |
Physical |
|
Dimensions |
303*216*89mm |
Weight |
2.0kg |
Operating Temperature |
0°C to 40°C |
Operating Humidity |
Relative humidity between 30% and 95% (non-condensing) |
Storage Temperature (without battery) |
-20°C to 55°C |
Storage Humidity (without battery) |
Up to 95% (non-condensing) |