MCA-VH1 can be used in respiratory management of after anesthesia operation, ambulance room, ICU, etc.
Specification:
Ventilator MCA-VH1 has volume control and pressure control working modes. The transduction from inspiration to expiration can be realized by setting the value of time, pressure, volume, and flow rate. Its preset TV range is 50-1500ml, Frequency is 2-60bpm, and flow rate is 10-120L/min. It can be applied to adult and pediatrics.
Ventilator MCA-VH1 can provide respiration modes include:A/C, SIMV, PCV, PSV, MMV, and SIGH. It also has PEEP support function. It can realize extra respiratory modes like CPAP, SIMV+PSV, inverse ration ventilation through combination of respiratory modes and setting parameters.
Ventilator MCA-VH1 provides the following alarms: TV upper/lower limit, peek pressure upper/lower limit, frequency upper/lower limit, MV upper/lower limit, oxygen concentration upper/lower limit, PEEP upper /lower limit, apnea, AC power cutoff. When power cutoff, the built-in battery will be drive the ventilator work.
Parameters:
A/C mode | Other parameters: | ||
1. Oxygen concentration: | 21~100% | 1. Oxygen pressure: | 50.75~79.75 psi |
2. PEEP pressure: | 0~20 cmH2O | 2. Noice | <65 dBA |
3. Tidal volume: | 50~1500 ml | 3. Environment: | |
4. Frequency: | 2~60 bpm | Temperature: | 5ºC~ 40ºC |
5. I:E ratio: | 3.0:1~1:3.0 | Relative humidity | ≤85%(25ºC) |
6. Waveform: | square waveform, speed-down waveform | Gas pressure: | 960~1060 cmH2O |
7. Inspiratory platform: | 0~50% | 4. Power: | A/C160~242V,50±1Hz |
8. Inspiration sensitivity: | -1~-20 or (OFF)cmH2O | 5. Power consumption: | 40VA |
9. SIGH: | 1.0~2.0 VT/50 | ||
Monitor parameters | |||
SIMV mode | 1.TV: | 0~9999 mL | |
1. Oxygen concentration: | 21~100% | 2. Respiratory frequency: | 0~100 bpm |
2. PEEP pressure: | 0~20 cmH2O | Machine-controlled frequency: | 0~100 bpm |
3. Tidal volume: | 50~1500 ml | Respiratory effort frequency: | 0~100 bpm |
4. SIMV frequency: | 2~35 bpm | 3. I:E ratio: | 3.0:1~1:9.9 |
5. Spontaneous support pressure: | 0~35 cmH2O | 4. Expiration end pressure: | 0~65 cmH2O |
6. Inspiration time: | 0.2~3.0 S | 5. Peak pressure | 0~68 cmH2O |
7. Inspiration sensitivity: | -1~ - 20 cmH2O | 6. MV: | 0~99.9 L/min |
8. Flow rate ascending speed: | 10~400 L/min/S | 7.Machine-controlled MV: | 0~99.9 L/min |
9. Expiration sensitivity: | 10 -75% | Respiratory effort MV: | 0~99.9 L/min |
10. Max. time of inspiration time: | 1.0~6.0 S | 7. Oxygen concentration | 21~100 % |
11. Apnea time: | 3~12 S | 8. inspiratory pressure: | 0~65 cmH2O |
PCV mode | Alarm parameters | ||
1. Oxygen concentration: | 21~100% | 1. TV upper limit: | 100~2000 ml |
2. PEEP pressure: | 0~20 cmH2O | 2. TV lower limit: | 20~2000 ml |
3. Frequency | 2~60 bpm | 3. Frequency upper limit: | 5~100 bpm |
4. Inspiratory pressure: | 0~35 cmH2O | 4. Frequency lower limit: | 2~60 bpm |
5. I:E ratio: | 3.0:1~1:3.0 | 5. Peak pressure upper limit: | 5~30 cmH2O |
6. Inspiratory platform: | 0~50% | 6. Peak pressure lower limit: | 0~20 cmH2O |
7. Inspiration sensitivity: | -1~-20 or (OFF) cmH2O | 7.Oxygen concentration upper limit: | 30%~100% |
8. Flow rate ascending speed: | 10~400 L/min/S | 8.concentration lower limit: | 21%~60% |
9. Min. TV: | 50~1000 ml | 9. MV upper limit: | 4.0~40 L/min |
10. MV lower limit: | 2.0~20 L/min | ||
PSV mode | 11. Apnea alarm time: | 3~12 S | |
1.Oxygen concentration: | 21~100% | ||
2. PEEP : | 0~20 cmH2O | MMV mode | |
3. support pressure: | 0~35 cmH2O | 1. Oxygen concentration: | 21~100% |
4. Inspiration sensitivity: | -1~ -20 cmH2O | 2. PEEP : | 0~20 cmH2O |
5. Expiration sensitivity: | 10~75% | 3. Target TV: | 50~1500 ml |
6. Flow rate ascending speed: | 10~400 L/min/S | 4. Ventilation: | 0.5~40.0 L/min |
7. Min. TV: | 50~1000 ml | 5. Inspiration sensitivity: | -1~ -20 cmH2O |
8. Max. inspiration time: | 1.0~6.0 S | 6. Flow rate ascending speed: | 50~400 L/min/S |
9. Apnea time: | 3~12 S | 7. Expiration sensitivity: | 10~75% |
10. spare frequency | 5~35 bpm | 8. Max. inspiration time: |
1.0~6.0 S |