Customization: | Available |
---|---|
Classification: | Infusion Pump |
Type: | ICU Anesthesia |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Hospital Electric Portable Infusion Pump
Model: MCS0028
Features
1. 3.5 inch LCD screen for clear vision within 5 meters
2. One-button night mode design for patient's more comfortable rest at night
3. Rate,Time-Rate,Volume-Rate,Time-Drop,Volume-Drop,Time-Volume to satisfy different clinical choices
4. 5 different brands of I.V sets can be calibrated and customized into the system
5. Numeric but easy cooperation more compatible with clinical use in ICU, NICU, OT etc.
Specifications
Model No.Parameters
|
MCS0028 Infusion Pump
|
|
Infusion Principle
|
Fingertip Peristaltic Pump
|
|
I.V.Set Compatibility
|
Open System, match all brands of qualified PVC,TPE I.V.sets with external diameter of 3.8mm-4.2mm
|
|
Range of Flow Rate
|
0.1-1200ml/h
|
|
Infusion Mode
|
Rate, Rate-Time, Rate-Volume, Time-Volume, Drop-Volume, Drop-Time, Body Weight
|
|
Drug Library
|
20 drug list with drug code display
|
|
Purge/Bolus Rate
|
1-1200ml/h,default 800ml/h, step by 1ml/h
|
|
Single Bolus Volume
|
1.0-10ml Adjustable, default 3 ml
|
|
Time Preset
|
00: 01~99: 59 (Hour: Min)
|
|
Volume Range
|
1~9999ml
|
|
Drop Range
|
1-400d/min, step by 1 drop
|
|
Accuracy
|
±5%
|
|
Total Volume Infused
|
0-9999 ml
|
|
Occlusion Pressure
|
High
|
40 KPa±20KPa
|
Middle
|
60 KPa±20KPa
|
|
Low
|
100KPa±20KPa
|
|
Air in line Detection
|
Ultrasonic wave
|
|
Visible & Audible Alarms |
Air in line, Down Stream Occlusion, Door Open, VTBI Completion, Near Completion, Low Battery, Battery Exhausted, Low Temperature,
Motor Malfunction, A.C Disconnection, A.C connection, I.V set dislocation, MPU Error, Battery Charging, Battery Charging Completion, Circuit Malfunction |
|
K.V.O Rate
|
1ml/h-5ml/h, default value 1ml/h, can be programmed by user step 0.1ml/h
|
|
Internal Battery
|
Lithium battery, 11.1/2000mAh, more than 4 hours working back up
|
|
Power Consumption
|
30VA
|
|
Power
|
AC 100V-240V 50HZ/60HZ
|
|
Classification
|
Class II, Type CF, IPX4
|
|
Dimension & Weight
|
13×17.5×23 cm;2Kg
|
|
Optional Function
|
Ambulance DC 12V
|