Customization: | Available |
---|---|
Customized: | Non-Customized |
Certification: | CE |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
1. Microprocessor based servo controlled temperature system
2. Air temperature, skin temperature, timer and heating power separately displayed by LED
3.Skin temperature sensor failure protect function avoid over temperature
4. Integral timer to record therapy time
5.Alarm functions: Power failure; Over temperature; Temperature deviation; Temperature sensor failure; Fan failure;
6. Temperature deviation and over temperature could be amended in front panel directly
7. Slience/reset button for alarms
8. >37 ºC temperature setting
9. A keypad lock prevent inadvertent changes the setting
10. A second thermal cut-off function for more safety
11. Humidity is adjustable continuously
12. The infant bed inclination angle is adjustable continuously
13. Trough for pulling and pushing, which is convenient to add water and wash
14. Fan with low noise and long life-span
15. Six operating windows, two sides access operating windows can be swiveled
16. Four drawers for easy storage
17. Heater is electrically cut-out if air temperature inside incubator exceed 40 ºC
18.RS232 Connector
19.Monitor Tray
20. Control part can be pull out in whole,convenient for clean and maintain
21.Oxygen inlet
1. Power supply: AC110/220V 60/50Hz
2. Power input: ≤ 650VA
3 Optional parts:
a. Environment temperature: 20~30ºC
b. Environment relative humidity: 30%~75%RH
c. Environment air velocity of flow: <0.3m/s
d.Ambient air pressure: 700 hPa -1060 hPa
4.Air Temperature control range: 25ºC-37ºC, 37.1ºC-38ºC
5. Skin temperature control range: 32ºC~37ºC, 37.1ºC~38ºC
6. Skin temperature sensor accuracy: +/-0.3ºC
7. Temperature fluctuation: +/-0.5ºC
8. Uniformity of mattress temperature: ≤ 0.8ºC
9. Humidity accuracy: ≤ 10%RH
10. Warm-up time: ≤ 45min
11. Internal noise level: ≤ 55dB
12. Infant bed tilt angle: +/-12 degree
13. Mattress size: 65cm(L)X 37cm(W)
14. Trough capacity: 1200ml
15. Air filter: 0.5um
16.Max CO2-concentration is 0.4%
17.Alarm:
a.The over temp of air temp : 38ºC
b.The second thermal cut-out temperature: 40ºC
c.The alarm for deviation: - Air temp: ±3.0ºC - Skin temp : ±1.0ºC
d.Alarm for Sensors:
When the skin (or air) temperature sensor is shut off or short circuit,cut off the heater,give out alarm with sound and light .
e.Alarm for power failure
The power fails,give out alarm with sound and light.
f.Alarm for fan failure
The fan stops working or has fault, give out alarm with sound and light and cut off the heater power.